| Accuracy: |
+/-
10% |
| Range: |
0.1
to 20.0 ppm (0.1-PPM resolution), or 0.01 to 2.00 ppm (0.01-PPM
resolution) |
| Sensor
Type: |
Electrochemical
Cell |
| Sensor
Life: |
Typically
1.5 to 2 years |
| Operating
Temperature: |
-25°C
to 55°C (-13°F to 131°F) |
| Humidity
Range: |
0
to 99% non-condensing for display; 0-95% non condensing for sensor |
| Relays: |
3
assignable relay outputs (10A-120V each) |
| Relay
Functions: |
Configurable
for normal/fail safe and latching/non-latching |
| Display: |
Highly
visible digital readout (even in sunlight) |
| Alarm: |
Optional
Audio - Piezoelectric buzzer, 85 dB @ 1ft |
| Trouble
Function: |
Indicates
loss of signal in sensor wire |
| Sensor
Cable length: |
Up
to 1000-ft |
| Power
Supply: |
120V/60Hz |
| Enclosure: |
High
impact Noryl |
| Dimensions: |
2.8”
W x 3.6” L x 2.3” D (70mm x 90mm x 58mm)
|
| Weight: |
1.2
kg (2.5 lbs) |